dc.contributor.authorDrobetz, Herwigen_US
dc.contributor.authorSchueller, Michaelen_US
dc.contributor.authorTschegg, Elmar Karlen_US
dc.contributor.authorHeal, Clareen_US
dc.contributor.authorRedl, Heinzen_US
dc.contributor.authorMuller, Reinholden_US
dc.date.accessioned2021-08-31T00:31:14Z-
dc.date.available2021-08-31T00:31:14Z-
dc.date.issued2011-01-
dc.identifier.citationDrobetz, H., Schueller, M., Tschegg, E. K., Heal, C., Redl, H., & Muller, R. (2011). Influence of screw diameter and number on reduction loss after plating of distal radius fractures. ANZ JOURNAL OF SURGERY, 81(1–2), 46–51. https://doi.org/10.1111/j.1445-2197.2010.05479.xen_US
dc.identifier.urihttp://dora.health.qld.gov.au/qldresearchjspui/handle/1/1490-
dc.description.abstractThe current options for plate-screw combinations in volar locking distal radius plates used for the treatment of distal radius fractures are either plates with a single distal screw row or plates with multiple distal screw rows. Additionally, the screws themselves may have either fixed angle locking or polyaxial locking mechanisms. To date, there is no evidence or consensus regarding the optimal plate-screw combination. The aim of this study was to assess the biomechanical behaviour of different plate-screw combinations with respect to total distal screw number, number of distal screw rows and screw projection surface area of the most distal row. Biomechanical study to assess six different plating configurations in five different volar locking plate models in a Sawbone distal radius fracture model. The specimens were loaded with 800 Newton loads for 2.000 cycles at 1 Hz. After cyclic loading, load-to-failure testing was performed.   With cyclical testing, there was a significant and positive correlation between rigidity and a greater projection area of the most distal screws. Dorsal tilting was significantly more pronounced in plate models with a lesser projection area of the most distal screws and a smaller number of distal screws. With load-to-failure testing, there was a significant increase in rigidity with increasing screw projection area of the most distal row and total number of distal screws. Additional distal screw rows in volar locking distal radius plates might not add substantially to resistance against loss of reduction in the post-operative period.en_US
dc.language.isoenen_US
dc.publisherWILEY-BLACKWELL PUBLISHING, INCen_US
dc.relation.ispartofANZ journal of surgeryen_US
dc.subjectbiomechanicalen_US
dc.subjectdistal radius fractureen_US
dc.subjectfixed angleen_US
dc.subjectlocking screwen_US
dc.subjectvolar locking plateen_US
dc.titleInfluence of screw diameter and number on reduction loss after plating of distal radius fracturesen_US
dc.typeArticleen_US
dc.identifier.doi10.1111/j.1445-2197.2010.05479.x-
item.languageiso639-1en-
item.openairecristypehttp://purl.org/coar/resource_type/c_18cf-
item.cerifentitytypePublications-
item.openairetypeArticle-
item.grantfulltextnone-
item.fulltextNo Fulltext-
Appears in Sites:Mackay HHS Publications
Show simple item record

Page view(s)

78
checked on Jan 30, 2025

Google ScholarTM

Check

Altmetric


Items in DORA are protected by copyright, with all rights reserved, unless otherwise indicated.